Energy Spectrum Extraction and Optimal Imaging via Dual-Energy Material Decomposition
Zhao, Wan, Zhang, Zhang, Niu · physics.med-ph,physics.optics · 2015-12-23 · 原文
Inferior soft-tissue contrast resolution is a major limitation of current CT scanners. The aim of the study is to improve the contrast resolution of CT scanners using dual-energy acquisition. Based on dual-energy material decomposition, the proposed method starts with extracting the outgoing energy spectrum by polychromatic forward projecting the material-selective images. The extracted spectrum is then reweighted to boost the soft-tissue contrast. A simulated water cylinder phantom with inserts that contain a series of six solutions of varying iodine concentration (range, 0-20 mg/mL) is used to evaluate the proposed method. Results show the root mean square error (RMSE) and mean energy difference between the extracted energy spectrum and the spectrum acquired using an energy-resolved photon counting detector(PCD), are 0.044 and 0.01 keV, respectively. Compared to the method using the standard energy-integrating detectors, dose normalized contrast-to-noise ratio (CNRD) for the proposed method are improved from 1 to 2.15 and from 1 to 1.88 for the 8 mg/mL and 16 mg/mL iodine concentration inserts, respectively. The results show CT image reconstructed using the proposed method is sup
讲义
讲义·推断 依据「原文」自动生成的结构化摘要(推断),非原文表述;以原文为准。
1. 人话版
Inferior soft-tissue contrast resolution is a major limitation of current CT scanners.
The aim of the study is to improve the contrast resolution of CT scanners using dual-energy acquisition.
2. 领域脉络
本文类目:physics.med-ph、physics.optics,属于其所在研究脉络的最新进展。
3. 机制拆解
Based on dual-energy material decomposition, the proposed method starts with extracting the outgoing energy spectrum by polychromatic forward projecting the material-selective images.
The extracted spectrum is then reweighted to boost the soft-tissue contrast.
4. 证据与数字
A simulated water cylinder phantom with inserts that contain a series of six solutions of varying iodine concentration (range, 0-20 mg/mL) is used to evaluate the proposed method.
Results show the root mean square error (RMSE) and mean energy difference between the extracted energy spectrum and the spectrum acquired using an energy-resolved photon counting detector(PCD), are 0.044 and 0.01 keV, respectively.
Compared to the method using the standard energy-integrating detectors, dose normalized contrast-to-noise ratio (CNRD) for the proposed method are improved from 1 to 2.15 and from 1 to 1.88 for the 8 mg/mL and 16 mg/mL iodine concentration inserts, respectively.